Validation of an online replanning technique for prostate adaptive radiotherapy

  • 1. Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, WI (United States)

Description

We have previously developed an online adaptive replanning technique to rapidly adapt the original plan according to daily CT. This paper reports the quality assurance (QA) developments in its clinical implementation for prostate cancer patients. A series of pre-clinical validation tests were carried out to verify the overall accuracy and consistency of the online replanning procedure. These tests include (a) phantom measurements of 22 individual patient adaptive plans to verify their accuracy and deliverability and (b) efficiency and applicability of the online replanning process. A four-step QA procedure was established to ensure the safe and accurate delivery of an adaptive plan, including (1) offline phantom measurement of the original plan, (2) online independent monitor unit (MU) calculation for a redundancy check, (3) online verification of plan-data transfer using an in-house software and (4) offline validation of actually delivered beam parameters. The pre-clinical validations demonstrate that the newly implemented online replanning technique is dosimetrically accurate and practically efficient. The four-step QA procedure is capable of identifying possible errors in the process of online adaptive radiotherapy and to ensure the safe and accurate delivery of the adaptive plans. Based on the success of this work, the online replanning technique has been used in the clinic to correct for interfractional changes during the prostate radiation therapy.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/56/12/013

Additional details

Identifiers

DOI
10.1088/0031-9155/56/12/013;
PII
S0031-9155(11)70944-7;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
56
Journal Issue
12
Journal Page Range
p. 3659-3668
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43022799
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; COMPUTER CODES; ERRORS; IMPLEMENTATION; MONITORS; NEOPLASMS; PATIENTS; PHANTOMS; PROSTATE; QUALITY ASSURANCE; RADIOTHERAPY; VALIDATION; VERIFICATION
Descriptors DEC
BODY; DISEASES; GLANDS; MALE GENITALS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; STRUCTURAL MODELS; TESTING; THERAPY