Validation of an online replanning technique for prostate adaptive radiotherapy
- 1. Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, WI (United States)
Description
We have previously developed an online adaptive replanning technique to rapidly adapt the original plan according to daily CT. This paper reports the quality assurance (QA) developments in its clinical implementation for prostate cancer patients. A series of pre-clinical validation tests were carried out to verify the overall accuracy and consistency of the online replanning procedure. These tests include (a) phantom measurements of 22 individual patient adaptive plans to verify their accuracy and deliverability and (b) efficiency and applicability of the online replanning process. A four-step QA procedure was established to ensure the safe and accurate delivery of an adaptive plan, including (1) offline phantom measurement of the original plan, (2) online independent monitor unit (MU) calculation for a redundancy check, (3) online verification of plan-data transfer using an in-house software and (4) offline validation of actually delivered beam parameters. The pre-clinical validations demonstrate that the newly implemented online replanning technique is dosimetrically accurate and practically efficient. The four-step QA procedure is capable of identifying possible errors in the process of online adaptive radiotherapy and to ensure the safe and accurate delivery of the adaptive plans. Based on the success of this work, the online replanning technique has been used in the clinic to correct for interfractional changes during the prostate radiation therapy.
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/56/12/013Additional details
Identifiers
- DOI
- 10.1088/0031-9155/56/12/013;
- PII
- S0031-9155(11)70944-7;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 56
- Journal Issue
- 12
- Journal Page Range
- p. 3659-3668
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43022799
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; COMPUTER CODES; ERRORS; IMPLEMENTATION; MONITORS; NEOPLASMS; PATIENTS; PHANTOMS; PROSTATE; QUALITY ASSURANCE; RADIOTHERAPY; VALIDATION; VERIFICATION
- Descriptors DEC
- BODY; DISEASES; GLANDS; MALE GENITALS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; STRUCTURAL MODELS; TESTING; THERAPY