Published October 2006 | Version v1
Journal article

The use of a commercial QA device for daily output check of a helical tomotherapy unit

  • 1. Department of Therapeutic Radiology-Radiation Oncology, University of Minnesota, Mayo Mail Code 494 420 Delaware Street S.E., Minneapolis, Minnesota 55455 (United States)

Description

Helical tomotherapy radiation therapy units, due to their particular design and differences from a traditional linear accelerator, require different procedures by which to perform routine quality assurance (QA). One of the principal QA tasks that should be performed daily on any radiation therapy equipment is the output constancy check. The daily output check on a Hi-Art TomoTherapy unit is commonly performed utilizing ionization chambers placed inside a solid water phantom. This provides a good check of output at one point, but does not give any information on either energy or symmetry of the beam, unless more than one point is measured. This also has the added disadvantage that it has to be done by the physics staff. To address these issues, and to simplify the process, such that it can be performed by radiation therapists, we investigated the use of a commercially available daily QA device to perform this task. The use of this device simplifies the task of daily output constancy checks and eliminates the need for continued physics involvement. This device can also be used to monitor the constancy of beam energy and cone profile and can potentially be used to detect gross errors in the couch movement or laser alignment

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
33
Journal Issue
10
Journal Page Range
p. 3680-3682
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38026578
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BEAMS; COMPUTERIZED TOMOGRAPHY; ERRORS; IONIZATION CHAMBERS; LINEAR ACCELERATORS; PHANTOMS; QUALITY ASSURANCE; RADIOTHERAPY
Descriptors DEC
ACCELERATORS; DIAGNOSTIC TECHNIQUES; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIOLOGY; STRUCTURAL MODELS; THERAPY; TOMOGRAPHY

Optional Information

Notes
(c) 2006 American Association of Physicists in Medicine