Published 1977 | Version v1
Report Open

Establishment of a precision long counter as a secondary standard for fast neutron flux density measurement

  • 1. Bhabha Atomic Research Centre, Bombay (India). Electronics Div.

Description

A precision long counter based on the design by De Pangher and Nichols has been set up as a standard for fast neutron flux density measurements. Results of measurements carried out in this laboratory to determine the characteristics of this detector, using a wide range of neutron standards such as Sb/Bw (γ,n); 252Cf; Ra/Be (γ,n); Ra/Be (α,n); Am/Be (α,n) and Pu/Be (α,n) are given. The efficiency of the detector is found to be constant within +- 5% in the energy range of 21 KeV to 4.5 MeV. A comparison is made with the results obtained by earlier workers. (author)

Availability note (English)

MF available from INIS under the Report Number.

Files

9374334.pdf

Files (224.8 kB)

Name Size Download all
md5:7c52a4e34de24cb52f2022779c5b6d12
224.8 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
5 p.
Report number
BARC--916