A comparison research of the energy response for seven commercially available mammography dosimeters

  • 1. Chinese Center for Disease Control and Prevention, Beijing (China). National Inst. for Radiological Protection

Description

Objective: To evaluate the least energy dependent response dosimeters in the comparison research. Methods: According to GB/T 19629-2005, the energy response of seven commercially available mammography dosimeters was determined for ten standard radiation qualities in the range 20-40 kV at the SSDL using a reference-class dosimeter. Results: The error range of correction factors of energy response, three type chambers are 3.0%, 3.3% and 3.3% respectively; three type semiconductor detectors are 3.8%, 36% and 43% and one type thermoluminescent dosimeter is 29%. Conclusion: Radcal 9095 dosimeter have the least energy response. Other type chambers have better energy response than semiconductors and TLD dosimeters. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Health
Journal Volume
15
Journal Issue
2
Journal Page Range
p. 151-152, 153
ISSN
1004-714X

Optional Information

Notes
1 fig., 3 tabs., 3 refs.