A comparison research of the energy response for seven commercially available mammography dosimeters
Creators
- 1. Chinese Center for Disease Control and Prevention, Beijing (China). National Inst. for Radiological Protection
Description
Objective: To evaluate the least energy dependent response dosimeters in the comparison research. Methods: According to GB/T 19629-2005, the energy response of seven commercially available mammography dosimeters was determined for ten standard radiation qualities in the range 20-40 kV at the SSDL using a reference-class dosimeter. Results: The error range of correction factors of energy response, three type chambers are 3.0%, 3.3% and 3.3% respectively; three type semiconductor detectors are 3.8%, 36% and 43% and one type thermoluminescent dosimeter is 29%. Conclusion: Radcal 9095 dosimeter have the least energy response. Other type chambers have better energy response than semiconductors and TLD dosimeters. (authors)
Additional details
Publishing Information
- Journal Title
- Chinese Journal of Radiological Health
- Journal Volume
- 15
- Journal Issue
- 2
- Journal Page Range
- p. 151-152, 153
- ISSN
- 1004-714X
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 38091188
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; COMPARATIVE EVALUATIONS; CORRECTIONS; DOSEMETERS; ENERGY DEPENDENCE; ERRORS; IONIZATION CHAMBERS; KEV RANGE 10-100; MAMMARY GLANDS; RADIATION QUALITY; SEMICONDUCTOR DETECTORS; SSDL; STANDARDS; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DOSEMETERS; ENERGY RANGE; EVALUATION; GLANDS; KEV RANGE; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIATION DETECTORS; RADIOLOGY
Optional Information
- Notes
- 1 fig., 3 tabs., 3 refs.