Induction pick-up of placement of electron beam centre of gravity

Description

For diagnostics of SILUND-20 electron beam induction pick-ups of placement of centre of gravity have been developed. The detector is a ferrite ring with four winding, confined into the metallic screen. The signals are fed to the electron blocks in CAMAC standard and further on, into computer, where the coordinates of center of gravity and the beam charge are calculated. The detector construction is described. The estimation of its precision is given. The results of on-line tests on the measurements of pick-up characteristics are presented: - resolution on X and Y in the circle with the radius of 3 cm is not less than 1 mm; - precision of the charge measurement amounts to 0.5%; - leading and travelling edges are not longer than 3 ns

Availability note (English)

MF available from INIS under the Report Number.

Files

14742656.pdf

Files (157.5 kB)

Name Size Download all
md5:c6f0839299ba26e960f34b9275beb14e
157.5 kB Preview Download

System files (13.0 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Индукционный измеритель положения центра электронного пучка

Publishing Information

Imprint Pagination
4 p.
Report number
JINR--13-82-221

INIS

Country of Publication
USSR
Country of Input or Organization
USSR
INIS RN
14742656
Subject category
S43: PARTICLE ACCELERATORS;
Descriptors DEI
ACCURACY; BEAM POSITION; CALIBRATION; ELECTRON BEAMS; MAGNETOINDUCTION SENSORS; PERFORMANCE
Descriptors DEC
BEAM MONITORS; BEAMS; LEPTON BEAMS; MEASURING INSTRUMENTS; PARTICLE BEAMS

Optional Information

Notes
2 refs.; 3 figs.