Published November 2019 | Version v1
Journal article

A study in using MICROMEGAS to improve particle identification with the TAMU-MDM focal plane detector

  • 1. National Institute of Physics and Nuclear Engineering, Bucharest-Magurele, RO 077125 (Romania)
  • 2. Cyclotron Institute, Texas A & M University, College Station, TX, 77843-3366 (United States)
  • 3. IRFU, CEA, Université Paris-Saclay, F-91191, Gif-sur-Yvette (France)
  • 4. CERN (EP-DT-ED), Geneva (Switzerland)

Description

A MICROMEGAS detection amplifier has been incorporated into the design of the TAMU-MDM focal plane detector with the purpose of improving the energy resolution and thus, the particle identification. Beam tests showed a factor of 2 improvement over the original design, from 10%–12% to 4%–6%, for ions with A≤40 at E/A ∼ 10–20 MeV.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2019.162461

Additional details

Identifiers

DOI
10.1016/j.nima.2019.162461;
PII
S0168900219310198;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
943
Journal Page Range
vp.
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
56005651
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
DESIGN; DETECTION; ENERGY RESOLUTION; IONS; PARTICLE IDENTIFICATION
Descriptors DEC
CHARGED PARTICLES; RESOLUTION

Optional Information

Copyright