Accuracy and detection limits for bioassay measurements in radiation protection. Statistical considerations
Description
This report provides statistical concepts and formulas for defining minimum detectable amount (MDA), bias and precision of sample analytical measurements of radioactivity for radiobioassay purposes. The defined statistical quantities and accuracy criteria were developed for use in standard performance criteria for radiobioassay, but are also useful in intralaboratory quality assurance programs. This report also includes a literature review and analysis of accuracy needs and accuracy recommendations of national and international scientific organizations for radiation or radioactivity measurements used for radiation protection purposes. Computer programs are also included for calculating the probabilities of passing or failing multiple analytical tests for different acceptable ranges of bias and precision
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS, PC A06/MF A01 - GPO as TI86901352.
Files
Additional details
Publishing Information
- Imprint Pagination
- 102 p.
- Report number
- NUREG--1156
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 17076312
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; RADIOACTIVITY; RADIOASSAY; RADIOISOTOPES; RADIOMETRIC ANALYSIS; STATISTICS
- Descriptors DEC
- CHEMICAL ANALYSIS; ISOTOPES; MATHEMATICS; QUANTITATIVE CHEMICAL ANALYSIS