Published November 1995 | Version v1
Report Open

International whole body counter intercomparison based on BOMAB phantom simulating 4 years old child

  • 1. ENEA, Bologna (Italy). Dipt. Ambiente

Description

In April 1993 a whole body counter intercomparison campaign, The 1993 Intercomparison/Intercalibration, started. The campaign has been organized by The Canadian National Reference Centre for In-Vivo Monitoring of Radiation Protection Bureau, Health Canada and The United States Department of Energy and it was based on measurements on a BOMAB type phantom simulating a 4 years old child. The phantom was filled with radioactive tissue substitute resin and an unknown quantity of radioactivity. Each facility was asked to determine the identity and amount of the radionuclide(s), knowing that the specific activity in the 10 BOMAB's sections was the same. Each facility was also asked to calculate the minimum detectable activity of all the radionuclides detected in the phantom. 35 Facilities from 20 different Countries took part in the initiative. The Institute for Radiation Protection of the Environment Department of ENEA (ENEA AMB IRP) represented Italy. Intercomparison results supplied by ENEA AMB IRP as radionuclides identification, activity data and associated precision, minimum detectable activity levels, can be considered satisfactory and comparable with results supplied by similar-facilities

Availability note (English)

MF available from INIS under the Report Number.

Files

28013120.pdf

Files (485.4 kB)

Name Size Download all
md5:06068b5332ee2f24d92f187021a4127b
485.4 kB Preview Download

Additional details

Additional titles

Original title (Italian)
Partecipazione dell'ENEA alla campagna internazionale di interconfronto wholebody counter con fantoccio simulante un bambino di 4 anni

Publishing Information

Imprint Pagination
23 p.
ISSN
1120-5555
Report number
ENEA-RT-AMB--95-17