Published August 1, 2003 | Version v1
Journal article

Results of long-term performance and subsequent laboratory tests of RPCs of the L3 forward-backward muon spectrometer

Description

The RPC chambers in the L3 forward-backward muon spectrometer have been operational from 1994 to the end of LEP running. After dismantling of the L3 detector some of the RPCs modules have been transported to Napoli where their performance has been re-measured with cosmic rays. Results of long-term performance in the LEP environment and of laboratory tests with cosmic rays will be presented. One of the 20 tested chambers, the one with the worst performances, was opened. The chamber was inspected with microscope and also a chemical analysis of the oil coating was performed. We will report in detail about this analysis

Additional details

Identifiers

PII
S0168900203013433;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
508
Journal Issue
1-2
Journal Page Range
p. 159-165
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
6. international workshop on resistive plate chambers and related detectors
Dates
26-27 Nov 2001
Place
Coimbra (Portugal)

INIS

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.