Published January 2007 | Version v1
Journal article

Multiple-estimate Monte Carlo calculation of the dose rate constant for a cesium-131 interstitial brachytherapy seed

  • 1. Radioisotopes Program, Pacific Northwest National Laboratory, Richland, Washington, 99354 (United States)

Description

The purpose of this study was to calculate a more accurate dose rate constant for the 131Cs (model CS-1, IsoRay Medical, Inc., Richland, WA) interstitial brachytherapy seed. Previous measurements of the dose rate constant for this seed have been reported by others with incongruity. Recent direct measurements by thermoluminescence dosimetry and by gamma-ray spectroscopy were about 15% greater than earlier thermoluminescence dosimetry measurements. Therefore, we set about to calculate independent values by a Monte Carlo approach that combined three estimates as a consistency check, and to quantify the computational uncertainty. The calculated dose rate constant for the 131Cs seed was 1.040 cGy h-1 U-1 for an ionization chamber model and 1.032 cGy h-1 U-1 for a circular ring model. A formal value of 2.2% uncertainty was calculated for both values. The range of our multiestimate values were from 1.032 to 1.061 cGy h-1 U-1. We also modeled three 125I seeds with known dose rate constants to test the accuracy of this study's approach

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
34
Journal Issue
1
Journal Page Range
p. 49-54
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2007 American Association of Physicists in Medicine