Published 1984 | Version v1
Report Open

Monte-Carlo calculation of the response function of the big scintillation detector for neutron capture detection

Description

Technique for calculation of the response function of a large liquid scintillation detector of radiation capture events is described. The GAMBAK and CYLIND programs permittinq to calculate γ-radiation energy losses in the sample and scintillator under the conditions of real geometry and a complex multicascade source usino the Monte-Carlo method are described in brief. The results of the calculations are compared with experimental amplitude detector spectra. The developed technique permits to calculate efficiency of a real detector at zero threshold as well as to obtain the correction for the ratio of the spectrum lying above the threshold. This technique permits also to determine corrections to self-absorption of capture γ-rays in samples

Availability note (English)

MF available from INIS under the Report Number.

Files

17023086.pdf

Files (456.7 kB)

Name Size Download all
md5:003825ef559deb0c27e76427228ec181
456.7 kB Preview Download

System files (37.8 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Расчет функции отклика большого сцинтилляционного детектора событий захвата нейтронов методом Монте-Карло

Publishing Information

Imprint Pagination
14 p.
Report number
FEI--1604

Optional Information

Notes
13 refs.; 5 figs.; 1 tab.