Published 1995 | Version v1
Report Open

Background characterization in a liquid scintillation spectrometer

Description

An alternate procedure for background count rate estimation in a liquid scintillation spectrometer is presented, which does not require to measure a blank with similar composition, volume and quench, to the problem sample. The procedure is based on a double linear parameterization which was obtained from a systematic study of the background observed with glass vials, in three different windows, 0 - 20 KeV, 0 - 800 KeV and 0 - 2 MeV, for volume between 2 and 20 mi of three commercial scintillators, Hisafe II, Ultima-Gold and Instagel, and quenching degree in the interval equivalent to 50% - 3% tritium efficiency. This procedure was tested with standard samples of 3H, and led to average discrepancies less than 10% for activity ≥0,6 Bq, against conventional methods for which the discrepancies are twice on average. (Author) 10 refs

Availability note (English)

Available from INIS in electronic form

Files

38094295.pdf

Files (1.4 MB)

Name Size Download all
md5:5f2ee44794af90160e73c9074248f881
1.4 MB Preview Download

Additional details

Additional titles

Original title (Spanish)
Caracterizacion del fondo de un espectrometro de centelleo liquido

Publishing Information

Imprint Pagination
32 p.
Report number
CIEMAT--758