Calculation of gamma-ray efficiency for 12.7 cm NE-213 organic scintillator detector
Description
The gamma-ray efficiency is reported for a 12.7 cm NE-213 organic scintillation detector. The efficiency was determined by Monte Carlo calculations. Results are given for incident gamma rays from 662 keV to 10 MeV. (Auth.)
Additional details
Identifiers
Publishing Information
- Journal Title
- Nuclear Instruments and Methods
- Journal Volume
- 129
- Journal Issue
- 2
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 619-620
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 7241140
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- EFFICIENCY; GAMMA DETECTION; GAMMA SPECTRA; KEV RANGE 100-1000; MEV RANGE 01-10; MONTE CARLO METHOD; ORGANIC COMPOUNDS; PLASTIC SCINTILLATION DETECTOR; RESPONSE FUNCTIONS
- Descriptors DEC
- ENERGY RANGE; FUNCTIONS; KEV RANGE; MEASURING INSTRUMENTS; MEV RANGE; RADIATION DETECTION; RADIATION DETECTORS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; SPECTRA