Description

A gamma ray spectrometer, with a 3'' X 3'' NaI(Tl) detector, with a moderator sphere has been utilised to measure the neutron fluence rate, with this value the H*(10) was estimated. When a neutron is captured by the hydrogen-based moderator, a 2.22 MeV prompt gamma ray is produced. In a multichannel analyser the net area under the 2.22 MeV photopeak is proportional to the total neutron fluence rate. The features of this system were determined by a Monte Carlo study that includes 3-, 5- and 10-inches diameter, water and polyethylene moderators and a 239Pu-Be source. The prompt gamma response was extended to monoenergetic neutron sources. To verify the response, a 239Pu-Be source in combination with a 10'' polyethylene sphere having a gamma-ray spectrometer with NaI(Tl) was utilised to estimate the neutron fluence rate and the H*(10). These results were compared with neutron fluence rate and H*(10) obtained using a Bonner sphere spectrometer and with the H*(10) measured using a neutron rem-meter. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncm055

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
126
Journal Issue
1-4
Journal Page Range
p. 265-268
ISSN
0144-8420

Conference

Title
10. International Symposium on Neutron Dosimetry - Progress in dosimetry of neutrons and light nuclei
Dates
12-16 Jun 2006
Place
Uppsala (Sweden)

Optional Information

Notes
21 refs