Miscellaneous Open

Patient dosimetry during chest radiography

  • 1. The VINCA Institute of Nuclear Sciences, POB. 522, 11001 Belgrade, (Serbia and Montenegro)

Description

Reasons for the variation in patient doses from chest radiography procedure were investigated by assessing entrance skin doses from kerma-area product measurements. Data were collected from seven x-ray tubes in five hospitals involving 259 adult patients. The third quartile value was 0.81 mGy compared to general reference level of 0.30 mGy. The applied tube potential was main contributor to patient dose variation. If department use at least 90 k Vp, the mean entrance surface dose would be reduced ut to factor six. Modification of departmental procedure is correct approach for dose reduction in diagnostic radiology. (author)

Availability note (English)

Available from INIS in electronic form; Also available from Institute of Nuclear Sciences VINCA, Library, PO Box 522, 11001 Belgrade, Serbia and Montenegro

Abstract (Serbian)

Nivoi izlozenosti pacijenta tokom radiografije pluca analizirani su merenjem proizvoda kerme i povrsine (KAP) za ukupno 259 odraslih pacijenata, u pet zdravstvenih ustanova. Koristeci izmerene vrednosti i podatke o radiografskoj tehnici izracunata je doza na povrsini koze pacijenta (ESD) za svakog pacijenta. Srednja vrednost KAP bila je 0.76 (0.22-1.88) Gy·cm2 a srednja vrednost ESD 0,78 (0.23-1.88) mGy. Za referntni nivo odabrana je vrednost treceg kvartila distribucije izmerenih vrednosti, sto iznosi 0.81 mGy. Osnovni uzrok varijacije pacijentnih doza nalazi se u primenjenoj radiografskoj tehnici: visokom naponu, rastojanju fokus-film i koriscenoj kombinaciji film-poja civacka folija. Smanjenje doza moze se postici korekcijom radioloske prakse: povecanjem napona i faktora pojacanja kombinacija film-pojacivacka folija u skladu sa telesnom masom pacijenta. (autor)

Files

37121852.pdf

Files (91.7 kB)

Name Size Download all
md5:81373dc7654e7cf917957b1d9023ae14
91.7 kB Preview Download

Additional details

Additional titles

Original title (Serbian)
Dozimetrijski aspekti radiografije pluca

Publishing Information

Publisher
Society for Electronics, Telecommunications, Computers, Automation and Nuclear Engineering. Section for Nuclear Techniques and Technology
Imprint Place
Beograd (Serbia and Montenegro)
Imprint Pagination
4 p.
Report number
INIS-RS--0225

Conference

Title
Society for Electronics, Telecommunications, Computers, Automation, and Nuclear Engineering
Original Conference Title
ETRAN 2004: 48. Konferencija za elektroniku, telekomunikacije, racunarstvo, automatiku i nuklearnu tehniku
Acronym
48. Conference - ETRAN 2004
Dates
6-10 Jun 2004
Place
Cacak (Serbia and Montenegro)

Optional Information

Notes
11 ref., 2 figs., 4 tabs.
Secondary number(s)
ISBN--86-80509-52-3