The patient exposure from chest examination in diagnostic radiology

  • 1. Institute of Public Health 'Prof. Dr. Iuliu Moldovan', 400349, Cluj-Napoca (Romania)

Description

The survey was carried out in the three representative units in Cluj city for chest examination (two hospitals and one ambulatory centre). The purpose of the study is to verify the most technical parameters of radiological system which influence the image quality and patient exposures. Values found for entrance surface dose (ESD) and effective doses are above diagnostic reference level in all three units included in our study, especially because the high value of mAs. Despite the same protocols applied for PhilipsDuodiagnost and Eltex X-ray system, the variation in doses is due to the inaccurate kVp. This high values of ratio patient effective dose / guidance level require an investigation and optimisation of X-ray system or protocols. (authors)

Files

36097742.pdf

Files (25.8 kB)

Name Size Download all
md5:2562def7ca20dc1477c305a3c514bca3
25.8 kB Preview Download
Part of:
IRPA Regional Congress on Radiation Protection in Central Europe

Additional details

Publishing Information

Publisher
Nuclear Regulatory Authority of the Slovak Republic
Imprint Place
Bratislava (Slovakia)
ISBN
80-88806-43-7
Imprint Title
IRPA Regional Congress on Radiation Protection in Central Europe
Imprint Pagination
573 p.
Journal Page Range
p. 1-5
Report number
INIS-SK--2005-030

Conference

Title
IRPA 2003
Dates
22-26 Sep 2003
Place
Bratislava (Slovakia)

Optional Information

Notes
3 refs., 7 tabs. Imprint:Also published on CD