Published April 2021 | Version v1
Journal article

Variation in dosimeter calibration factor (Ndw) over a period of 20 years

  • 1. Department of Radiation Oncology, Army Hospital Research and Referral, Delhi (India)
  • 2. Division of Clinical PET, Institute of Nuclear Medicine and Allied Sciences, Delhi (India)

Description

Long delays in renewal of calibration of secondary standards radiation dosimeters in radiation oncology centers due to the COVID19 pandemic have aroused concerns regarding accuracy in dose delivery to patients. The concerns are mainly due to the uncertainty in the absorbed dose to water calibration factor (NDW) over a period of time. In this study, the NDW factor for two ion chambers, thimble type (Farmer) and parallel plate type (Markus), used in most of the radiotherapy centers, were retrospectively reviewed for 20 years. The calibration on all occasions except once was carried out at the Secondary Standards Dosimetry Laboratory, Bhabha Atomic Research Centre, Mumbai. The change in the NDW factor over this period was < ±3%. We, therefore, believe that a dosimeter with no history of repairs showed reasonable stability in the NDW factor over a long period. (author)

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
46
Journal Issue
2
Journal Page Range
p. 125-127
CODEN
JMPHFE

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
52102936
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
ABSORBED DOSE RANGE; ABSORBED RADIATION DOSES; CALIBRATION; DOSEMETERS; DOSIMETRY; RADIATION DOSE DISTRIBUTIONS
Descriptors DEC
DOSES; MEASURING INSTRUMENTS; RADIATION DOSE RANGES; RADIATION DOSES