Published March 15, 2001 | Version v1
Journal article

A two-detector anticoincidence system for reduction of Compton scattered high energy γ-ray background in PIXE analysis

  • 1. Marburg Univ. (Germany, F.R.). Fachbereich Physik

Description

A Si(Li)/CsI(T1) two-detector anticoincidence system, applied for background suppression in X-ray spectra, is described. Since the high energy background in PIXE spectra is produced mainly by Compton scattering of γ-rays in the Si(Li) detector, the anticoincidence spectrometer can be used to improve the sensitivity of PIXE analysis for trace elements of higher atomic number (Mo, Cd, I). Details of the system and the performance of the spectrometer are presented. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods
Journal Volume
181
Journal Issue
1-3
Series
Nucl. Instrum. Methods.
Journal Page Range
25-29
ISSN
0029-554X

Conference

Title
2. International conference on particle induced X-ray emission and its analytical applications (PIXE-2).
Dates
9 - 12 Jun 1980.
Place
Lund, Sweden.