Published April 1, 2002
| Version v1
Journal article
Gray levels, dynamic range, contrast and DQE in imaging detectors
Creators
Description
A very important parameter which characterises an imaging detector is the maximum number (Gmax) of luminance levels, or gray levels, that it can discriminate. In this paper a physical model is proposed regarding the possibility of determining Gmax by computing the detectable minimum signal difference (the gray step) in various experimental conditions. As a consequence it is demonstrated that Gmax is different from the dynamic range of the detector. The possibility of determining the minimum perceptible contrast (MPC) and the equivalence of the performance analysis of an imaging detector in terms of gray scale, MPC and DQE are also demonstrated
Additional details
Identifiers
- PII
- S0168900201012621;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 481
- Journal Issue
- 1-3
- Journal Page Range
- p. 691-695
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34003486
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- IMAGE PROCESSING; IMAGES; PHYSICAL PROPERTIES; RADIATION DETECTION; SENSITIVITY
- Descriptors DEC
- DETECTION; PROCESSING
Optional Information
- Copyright
- Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.