Published April 1, 2002 | Version v1
Journal article

Gray levels, dynamic range, contrast and DQE in imaging detectors

Creators

Description

A very important parameter which characterises an imaging detector is the maximum number (Gmax) of luminance levels, or gray levels, that it can discriminate. In this paper a physical model is proposed regarding the possibility of determining Gmax by computing the detectable minimum signal difference (the gray step) in various experimental conditions. As a consequence it is demonstrated that Gmax is different from the dynamic range of the detector. The possibility of determining the minimum perceptible contrast (MPC) and the equivalence of the performance analysis of an imaging detector in terms of gray scale, MPC and DQE are also demonstrated

Additional details

Identifiers

PII
S0168900201012621;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
481
Journal Issue
1-3
Journal Page Range
p. 691-695
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34003486
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
IMAGE PROCESSING; IMAGES; PHYSICAL PROPERTIES; RADIATION DETECTION; SENSITIVITY
Descriptors DEC
DETECTION; PROCESSING