Published July 7, 2021 | Version v1
Journal article

Performance of a low gain avalanche detector in a medical linac and characterisation of the beam profile

  • 1. Department of Physics & Astronomy, University of Kansas, Lawrence, KS 66045 (United States)
  • 2. St. Luke's Hospital, Rathgar, Dublin 6 (Ireland)
  • 3. School of Physics, University College Dublin, Belfield, Dublin 4 (Ireland)

Description

Low gain avalanche detectors can measure charged particle fluences with high speed and spatial precision, and are a promising technology for radiation monitoring and dosimetry. A detector has been tested in a medical linac where single particles were observed with a time resolution of 50 ps. The integrated response is similar to a standard ionising chamber but with a spatial precision twenty times finer, and a temporal precision over 100 million times better, with the capability to measure the charge deposited by a single linac pulse. The unprecedented resolving power allows the structure of the ∼3 μs linac pulses to be viewed and the 350 ps sub-pulses in the train to be observed. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/ac0587

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
66
Journal Issue
13
Journal Page Range
[10 p.]
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
53063564
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; BEAM PROFILES; CHARGED PARTICLES; LINEAR ACCELERATORS; RADIATION MONITORING
Descriptors DEC
ACCELERATORS; MONITORING