Published March 5, 2009 | Version v1
Journal article

The KamLAND Full-Volume Calibration System

Description

We have successfully built and operated a source deployment system for the KamLAND detector. This system was used to position radioactive sources throughout the delicate 1-kton liquid scintillator volume, while meeting stringent material cleanliness, material compatibility, and safety requirements. The calibration data obtained with this device were used to fully characterize detector position and energy reconstruction biases. As a result, the uncertainty in the size of the detector fiducial volume was reduced by a factor of two. Prior to calibration with this system, the fiducial volume was the largest source of systematic uncertainty in measuring the number of antineutrinos detected by KamLAND. This paper describes the design, operation and performance of this unique calibration system.

Availability note (English)

Available from OSTI as DE00957064; PURL: https://www.osti.gov/servlets/purl/957064-JkxAjY/

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
4
Journal Page Range
p. 30
ISSN
1748-0221

Optional Information

Contract/Grant/Project number
AC02-05CH11231
Notes
Journal Publication Date: April 27, 2009
Funding organization
Nuclear Science Division (United States); Physics Division (United States)
Secondary number(s)
LBNL--1855E