Published January 1982 | Version v1
Journal article

Effect of window width and viewing distance in CT display

  • 1. Middlesex Hospital, London (UK). Medical School
  • 2. Queen Elizabeth Coll., London (UK). Dept. of Physics

Description

From this preliminary analysis, it may be seen that the noise-limited detectability of a 20 mm diameter object reaches a maximum at a viewing distance of about 4 m at low window widths, when a quarter of the scan field is displayed in a 74 mm diameter image. For the more usual situation in which the whole scan field is imaged, the optimum distance would be in the region of 2 m, considerably larger than the distance at which a radiologist normally views a patient scan. The alternative to increasing the viewing distance is to reduce the size of the image, and the author's hospital has now increased the number of images per film to nine. This should result in an increase in the detectability of low-contrast lesions as well as effecting a considerable reduction in film costs. (author)

Additional details

Publishing Information

Journal Title
Br. J. Radiol.
Journal Volume
55
Journal Issue
649
Series
Br. J. Radiol.
Journal Page Range
72-74
ISSN
0007-1285

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
13671723
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; COST; DISPLAY DEVICES; IMAGE PROCESSING; IMAGES; NOISE; PHOTOGRAPHIC FILMS; RESOLUTION; WINDOWS
Descriptors DEC
COMPUTER-GRAPHICS DEVICES; DIAGNOSTIC TECHNIQUES; OPENINGS; TOMOGRAPHY