Published May 2017 | Version v1
Miscellaneous

Review study on variables for performance improvement of BF3 combination detector

  • 1. Ulsan National Institute of Science and Technology, Ulsan (Korea, Republic of)

Description

In this study, variables affecting performance of BF3 neutron combination detector was suggested. Various variables including shield control to neutron backscattering, tube pressure, temperature, detector assembly are analyzed. From further study on neutron detector, neutron detection performance limitation by variables such as shield control to neutron backscattering, BF3 tube pressure, temperature, cross section, detector assembly would get more clearly and systematically solved. In this study, the fact that kind of shield could influence process of moderation or reflection of neutron or flux control or scattering and absorption block was checked. Also, the fact that pressure of tubes and number of tubes are directly connected to efficiency of detector could be checked. Temperature change of detector could change energy peak and pulse height distribution. Detector assembly, geometric arrangement and resolution of detector and dead time settings are another variables to change efficiency of detector and influence performance of detector. Utilizing these information, experiment design for detector performance improvement and investigation about efficiency improvement of combination detector would be accomplished. Ultimately, composition and development for enhanced accuracy of neutron detector are our next goal.

Part of:
Proceedings of the KNS 2017 Spring Meeting

Additional details

Publishing Information

Publisher
KNS
Imprint Place
Daejeon (Korea, Republic of)
Imprint Title
Proceedings of the KNS 2017 Spring Meeting
Imprint Pagination
[1 CD-ROM]
Journal Page Range
[4 p.]

Conference

Title
2017 Spring Meeting of the KNS
Dates
17-19 May 2017
Place
Jeju (Korea, Republic of)

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
49050434
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; BACKSCATTERING; CONTROL; CROSS SECTIONS; EFFICIENCY; NEUTRON DETECTORS; PERFORMANCE; SHIELDS
Descriptors DEC
MEASURING INSTRUMENTS; RADIATION DETECTORS; SCATTERING

Optional Information

Notes
6 refs, 2 figs, 1 tab