Stitching-angle measurable microscopic-interferometer: Surface-figure metrology tool for hard X-ray nanofocusing mirrors with large curvature
Creators
- 1. Japan Synchrotron Radiation Research Institute (JASRI)/SPring-8, 1-1-1 Kouto, Sayo-cho, Sayo-gun Hyogo 679-5198 (Japan)
- 2. PRESTO, Japan Science and Technology Agency, 4-1-8 Honcho Kawaguchi, Saitama (Japan)
- 3. Department of Precision Science and Technology, Graduate School of Engineering, Osaka University, 2-1 Yamada-oka, Suita, Osaka 565-0871 (Japan)
- 4. RIKEN/SPring-8, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148 (Japan)
Description
We propose an optical profilometer system for evaluating mirrors that are used for hard X-ray nanofocusing. A surface metrology system designed for these mirrors to focus to a size of 10 nm is required to ensure a precise measurement of a large curvature with a vertical accuracy of 1 nm and a lateral resolution less than 0.1 mm. The surface metrology system is based on a stitching interferometer that can determine stitching angles between adjacent local surface profiles. Local surface profiles and stitching angles are detected simultaneously by using two types of optical interferometers, a microscopic interferometer and a large-scale interferometer, respectively. We developed a nanofocusing mirror that has approximately the same surface curvature as a sub-10-nm focusing mirror, to examine the measurement accuracy of the developed surface metrology system. As a result of a performance assessment of the measurement instrument, we achieved measurement reproducibility and reliability of less than peak-to-valley height of 4 nm by measuring the focusing mirror. In addition, we achieved focusing performance of a full width at half maximum 27 nm at 7 keV using the mirror at SPring-8. We verified that the metrology system can be used to measure sub-10-nm focusing mirrors.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2009.11.073Additional details
Identifiers
- DOI
- 10.1016/j.nima.2009.11.073;
- PII
- S0168-9002(09)02288-8;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 616
- Journal Issue
- 2-3
- Journal Page Range
- p. 203-206
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- International workshop on X-ray mirror design, fabrication and metrology (IWXM)
- Dates
- 22-24 Sep 2009
- Place
- Osaka (Japan)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41077840
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; FOCUSING; HARD X RADIATION; HEIGHT; INTERFEROMETERS; INTERFEROMETRY; KEV RANGE 01-10; MIRRORS; PERFORMANCE; SPRING-8 STORAGE RING; SURFACES; WIDTH
- Descriptors DEC
- DIMENSIONS; ELECTROMAGNETIC RADIATION; ENERGY RANGE; IONIZING RADIATIONS; KEV RANGE; MEASURING INSTRUMENTS; RADIATION SOURCES; RADIATIONS; STORAGE RINGS; SYNCHROTRON RADIATION SOURCES; X RADIATION
Optional Information
- Copyright
- Copyright (c) 2009 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.