Published May 1, 2010 | Version v1
Journal article

Stitching-angle measurable microscopic-interferometer: Surface-figure metrology tool for hard X-ray nanofocusing mirrors with large curvature

  • 1. Japan Synchrotron Radiation Research Institute (JASRI)/SPring-8, 1-1-1 Kouto, Sayo-cho, Sayo-gun Hyogo 679-5198 (Japan)
  • 2. PRESTO, Japan Science and Technology Agency, 4-1-8 Honcho Kawaguchi, Saitama (Japan)
  • 3. Department of Precision Science and Technology, Graduate School of Engineering, Osaka University, 2-1 Yamada-oka, Suita, Osaka 565-0871 (Japan)
  • 4. RIKEN/SPring-8, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148 (Japan)

Description

We propose an optical profilometer system for evaluating mirrors that are used for hard X-ray nanofocusing. A surface metrology system designed for these mirrors to focus to a size of 10 nm is required to ensure a precise measurement of a large curvature with a vertical accuracy of 1 nm and a lateral resolution less than 0.1 mm. The surface metrology system is based on a stitching interferometer that can determine stitching angles between adjacent local surface profiles. Local surface profiles and stitching angles are detected simultaneously by using two types of optical interferometers, a microscopic interferometer and a large-scale interferometer, respectively. We developed a nanofocusing mirror that has approximately the same surface curvature as a sub-10-nm focusing mirror, to examine the measurement accuracy of the developed surface metrology system. As a result of a performance assessment of the measurement instrument, we achieved measurement reproducibility and reliability of less than peak-to-valley height of 4 nm by measuring the focusing mirror. In addition, we achieved focusing performance of a full width at half maximum 27 nm at 7 keV using the mirror at SPring-8. We verified that the metrology system can be used to measure sub-10-nm focusing mirrors.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2009.11.073

Additional details

Identifiers

DOI
10.1016/j.nima.2009.11.073;
PII
S0168-9002(09)02288-8;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
616
Journal Issue
2-3
Journal Page Range
p. 203-206
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
International workshop on X-ray mirror design, fabrication and metrology (IWXM)
Dates
22-24 Sep 2009
Place
Osaka (Japan)

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41077840
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; FOCUSING; HARD X RADIATION; HEIGHT; INTERFEROMETERS; INTERFEROMETRY; KEV RANGE 01-10; MIRRORS; PERFORMANCE; SPRING-8 STORAGE RING; SURFACES; WIDTH
Descriptors DEC
DIMENSIONS; ELECTROMAGNETIC RADIATION; ENERGY RANGE; IONIZING RADIATIONS; KEV RANGE; MEASURING INSTRUMENTS; RADIATION SOURCES; RADIATIONS; STORAGE RINGS; SYNCHROTRON RADIATION SOURCES; X RADIATION

Optional Information

Copyright
Copyright (c) 2009 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.