14: Accuracy of dose calculations of the Philips treatment planning system OSS for blocked fields
Creators
- 1. Wuerzburg Univ. (Germany, F.R.). Klinik und Poliklinik fuer Strahlentherapie
Description
The main features of the OSS model regarding blocked fields are discussed. For 60Co gamma rays, 8 MV and 16 MV X-rays results of the OSS are compared with measurements. The systematic underestimation of the dose by 3% to 10% is caused mainly by improper use of the equivalent square principle and use of one reference dose table for both, blocked and normal fields. Recommendations to improve the OSS program are given. 7 refs.; 7 figs
Additional details
Publishing Information
- Publisher
- North-Holland.
- Imprint Place
- Amsterdam (Netherlands)
- ISBN
- 0-444-70263-6
- Imprint Title
- The use of computers in radiation therapy
- Imprint Pagination
- 611 p.
- Journal Page Range
- p. 57-60.
Conference
- Title
- 9. International conference on the use of computers in radiation therapy.
- Dates
- 22-25 Jun 1987.
- Place
- Scheveningen (Netherlands).
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 20026816
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; COBALT 60; COMPARATIVE EVALUATIONS; COMPUTER CALCULATIONS; DOSIMETRY; MEV RANGE 10-100; PHOTON BEAMS; PLANNING; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- BEAMS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; COBALT ISOTOPES; ENERGY RANGE; EVALUATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTO; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEDICINE; MEV RANGE; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; RADIOISOTOPES; THERAPY; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- Imprint:refs.; figs.; tabs.