Published 2018 | Version v1
Book

A comparison study of two phoswich detectors used for lung monitoring with varying thickness of primary crystal

  • 1. Radiological Safety Division, Indira Gandhi Centre for Atomic Research, Kalpakkam (India)

Description

In-vivo monitoring of actinides like 239Pu, 241Am and natural uranium are carried out using Phoswich based lung monitoring system. The thickness of the primary detector used in Phoswich detector has to be optimized to get maximum efficiency with minimum background interference. Also the window material and its thickness as well as the inherent natural/fallout activity present in detector material has to be carefully looked into for better background reduction. Two phoswich detectors, of 3 mm thick and 15 mm thick primary NaI(Tl) crystals were compared for their performance using parameters such as efficiency, MDA, background interference etc and are discussed in this study. (author)

Part of:
Proceedings of the twenty first national symposium on radiation physics: book of abstracts cum souvenir

Additional details

Publishing Information

Publisher
Raja Ramanna Centre for Advanced Technology
Imprint Place
Indore (India)
Imprint Title
Proceedings of the twenty first national symposium on radiation physics: book of abstracts cum souvenir
Imprint Pagination
225 p.
Journal Page Range
p. 145

Conference

Title
21. DAE-BRNS national symposium on radiation physics
Acronym
NSRP-2018
Dates
5-7 Mar 2018
Place
Indore (India)

Optional Information

Notes
1 ref., 1 fig., 1 tab.