Published March 1985
| Version v1
Journal article
Investigation of the characteristics of multiwire proportional chambers with information pickup from the cathode
Creators
- 1. Inst. of High-Energy Physics, Serpukhov
Description
An accuracy of 37 μm in the determination of coordinates is obtained with 6-mm-wide cathode strips and an anode-cathode spacing of 4.8 mm. It is shown that the appearance of a self-quenching streamer discharge in the proportional chamber results in a deterioration of the spatial resolution of the instrument. The time resolution of the chambers is 16 nsec
Additional details
Publishing Information
- Journal Title
- Instrum. Exp. Tech. (Engl. Transl.)
- Journal Volume
- 27
- Journal Issue
- 5
- Series
- Instrum. Exp. Tech. (Engl. Transl.).
- Journal Page Range
- 1095-1098
- ISSN
- 0020-4412
- CODEN
- INETA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18006335
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; ANODES; CATHODES; MULTIWIRE PROPORTIONAL CHAMBER; PERFORMANCE; RADIATION STREAMING; SPATIAL RESOLUTION; TIME RESOLUTION
- Descriptors DEC
- ELECTRODES; MEASURING INSTRUMENTS; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RESOLUTION; TIMING PROPERTIES
Optional Information
- Notes
- Translated from Pribory i Tekhnika Eksperimenta, No. 5, pp. 47-49, September-October, 1984.. Cover-to-cover translation of Pribory i Tekhnika Ehksperimenta (USSR).