Published April 2017
| Version v1
Report
Open
Improvement of XRF and PXRF analysis using elemental sensitivity method
Creators
- 1. Atomic Energy Commission, Damascus (Syrian Arab Republic). Dept. of Chemistry
Description
AXIL-QXAS has been considered as an important program in qualitative and quantitative determination of elements by XRF. It is first where the possible use of Ar intensity in thickness determination of samples was studied for the analysis by the previous program. The analytical results showed the preferability of Ar-correlation method in comparison with X-ray absorption or determination of Compton to Rayleigh ratio methods. However, the range of metals (Fe, Cu, Zn) concentrations to achieve thick samples was 9.38-17.14 mg/g, while this range was 80-140 mg/g and a surface density more than 160 mg/cm2 for the X-ray absorption method (author).
Files
50071208.pdf
Files
(2.2 MB)
Name | Size | Download all |
---|---|---|
md5:63d7185d2a1800c971a2e4804c017377
|
2.2 MB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 19 p.
- Report number
- AECS-C/RSS--1167
INIS
- Country of Publication
- Syrian Arab Republic
- Country of Input or Organization
- Syrian Arab Republic
- INIS RN
- 50071208
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Numerical Data
- Quality check status
- Yes
- Descriptors DEI
- ABSORPTION; ARGON; ELEMENTS; EXPERIMENTAL DATA; PORTABLE EQUIPMENT; SENSITIVITY; THICKNESS; X-RAY FLUORESCENCE ANALYZERS;
- Descriptors DEC
- DATA; DIMENSIONS; ELEMENTS; EQUIPMENT; FLUIDS; GASES; INFORMATION; NONMETALS; NUMERICAL DATA; RARE GASES; SORPTION;
Optional Information
- Notes
- 10 refs. , 6 figs. , 2 tabs.