Published August 1996 | Version v1
Report Open

Simulation and radiation treatment in external radiotherapy

Creators

  • 1. Mevaterapia Medical Center, Buenos Aires (Argentina)

Description

It is well known that in order to obtain a uniform dose in the treated volume as defined in ICRU 50, there should be a 10% maximum difference between maximum and minimum dose values in treatment planning. Clinical target volume (CTV) should be related to external areas of body sections where tumour is located. These areas are important because different radiation beams enter through them. Therefore, verification of the planning target volume (PTV) through the external areas is highly significant. In this work we point out the importance of controlling that PTV is irradiated as planned considering some error sources usually found in radiotherapy practice with equipment that has been intensively used for a long time. Moreover, I think this experience will be helpful for those centers around the world where radiation treatment is carried out with reconditioned units. (author)

Files

28018209.pdf

Files (126.9 kB)

Name Size Download all
md5:f90b2876fa6ccf0a38f56453ed6edf88
126.9 kB Preview Download
Part of:
Radiation dose in radiotherapy from prescription to delivery

Additional details

Publishing Information

Imprint Title
Radiation dose in radiotherapy from prescription to delivery
Imprint Pagination
293 p.
Journal Page Range
p. 101-105.
ISSN
1011-4289
Report number
IAEA-TECDOC--896

Conference

Title
Radiation dose in radiotherapy from prescription to delivery.
Acronym
Interregional seminar on radiotherapy dosimetry
Dates
27-30 Aug 1994.
Place
Rio de Janeiro (Brazil).

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
28018209
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; EQUIPMENT; EXTERNAL IRRADIATION; HUMAN FACTORS; PLANNING; QUALITY CONTROL; RADIATION DOSES; RADIATION HAZARDS; RADIOTHERAPY; SIMULATION
Descriptors DEC
CONTROL; HAZARDS; HEALTH HAZARDS; IRRADIATION; MEDICINE; THERAPY

Optional Information