Published June 15, 1987
| Version v1
Journal article
On the correct choice of transformer ratio in low noise preamplifiers for large detector capacitances
- 1. Istituto Nazionale di Fisica Nucleare, Milan (Italy)
- 2. Politecnico di Milano (Italy). Dipartimento di Elettronica
- 3. Politecnico di Milano (Italy). Ist. di Fisica
- 4. Consiglio Nazionale delle Ricerche, Milan (Italy). Centro di Elettronica Quantistica e Strumentazione Elettronica
- 5. Pavia Univ. (Italy). Dipt. di Elettronica
Description
In the design of transformer coupled preamplifier for large detector capacitances the transformer ratio is customarily made equal to the value which meets the capacitive matching condition. If, however, parallel noise as well as series noise external to the amplifying device are present, the optimum choice of the transformer ratio may differ substantially from the previous one. The paper aims at investigating the optimum transformer ratio with a quite general configuration of noise sources, yet sticking to the model of ideal transformer. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res., Sect. A
- Journal Volume
- 257
- Journal Issue
- 2
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 331-338
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 19013956
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKGROUND NOISE; COUPLING; OPTIMIZATION; PREAMPLIFIERS; TRANSFORMERS
- Descriptors DEC
- AMPLIFIERS; ELECTRICAL EQUIPMENT; ELECTRONIC EQUIPMENT; EQUIPMENT; NOISE