Published August 21, 2001 | Version v1
Journal article

Extruded plastic counters with WLS fiber readout

Description

Extruded plastic scintillation counters with WLS fiber readout are described. For a 7 mm thick counter with 4.3 m long double-clad fibers spaced at 7 mm a light yield of 18.7 photoelectrons/MeV and a time resolution of 0.71 ns (σ) were obtained. A prototype photon veto module consisting of 10 layers of 7 mm thick grooved plastic slabs interleaved with 1 mm lead sheets was also tested, which yielded 122 photoelectrons per minimum ionizing particle and time resolution of 360 ps

Additional details

Identifiers

PII
S016890020100780X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
469
Journal Issue
3
Journal Page Range
p. 340-346
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.