Published 2014 | Version v1
Miscellaneous Open

Characterization of a free-air ionization chamber in direct X-ray beams as used in mammography

Description

At this work stability and characterization tests were undertaken on a Victoreen free-air ionization chamber, model 481. The tests were realized using direct X-ray beams as a contribution for its establishment as a primary standard system of the quantity air kerma. The characterization tests were: saturation curve, ion collection efficiency, polarity effect, response linearity with the air kerma rate and response linearity with the chamber volume variation. The ion collection efficiency allowed the determination of the ion recombination factor. Most of the test results showed agreement with the limits established by international standards. Furthermore, the air attenuation factors for the mammography beams with aluminum and molybdenum filters were obtained. The factors for photon transmission and scattering at the diaphragm edges were also determined for mammography beams with aluminum filter and for the standard beam with molybdenum filter. (author)

Files

45086004.pdf

Files (2.5 MB)

Name Size Download all
md5:9b892fb8644d6a2a7c7f8e8dce05a8db
2.5 MB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Caracterizacao de uma camara de ionizacao de ar-livre em feixes diretos de raios X utilizados em mamografia

Publishing Information

Imprint Pagination
94 p.
Report number
INIS-BR--14238