A portable tissue-equivalent proportional counter (TEPC)-based system was developed, which will be used as a reference standard in the neutron mixed field for radiation protection. In this paper, microdosimetry of the TEPC system was studied both in neutron (252Cf and 241Am-Be sources) and gamma (137Cs and 60Co sources) reference radiation fields. The measured neutron and gamma-dose equivalent rates were, respectively, compared with those of the reference ambient dose equivalent rate H*(10). And the measured microdosimetric spectra were compared with the Monte Carlo code FLUKA simulated results. All of the comparisons primarily agreed well. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nct221

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
158
Journal Issue
2
Journal Page Range
p. 246-250
ISSN
0144-8420

Optional Information

Notes
12 refs