MO-G-BRF-09: Investigating Magnetic Field Dose Effects in Mice: A Monte Carlo Study
- 1. UT MD Anderson Cancer Center, Houston, TX (United States)
Description
Purpose: In MRI-linac treatments, radiation dose distributions are affected by magnetic fields, especially at high-density/low-density interfaces. Radiobiological consequences of magnetic field dose effects are presently unknown; therefore, preclinical studies are needed to ensure the safe clinical use of MRI-linacs. This study investigates the optimal combination of beam energy and magnetic field strength needed for preclinical murine studies. Methods: The Monte Carlo code MCNP6 was used to simulate the effects of a magnetic field when irradiating a mouse-sized lung phantom with a 1.0cmx1.0cm photon beam. Magnetic field effects were examined using various beam energies (225kVp, 662keV[Cs-137], and 1.25MeV[Co-60]) and magnetic field strengths (0.75T, 1.5T, and 3T). The resulting dose distributions were compared to Monte Carlo results for humans with various field sizes and patient geometries using a 6MV/1.5T MRI-linac. Results: In human simulations, the addition of a 1.5T magnetic field caused an average dose increase of 49% (range:36%–60%) to lung at the soft tissue-to-lung interface and an average dose decrease of 30% (range:25%–36%) at the lung-to-soft tissue interface. In mouse simulations, the magnetic fields had no effect on the 225kVp dose distribution. The dose increases for the Cs-137 beam were 12%, 33%, and 49% for 0.75T, 1.5T, and 3.0T magnetic fields, respectively while the dose decreases were 7%, 23%, and 33%. For the Co-60 beam, the dose increases were 14%, 45%, and 41%, and the dose decreases were 18%, 35%, and 35%. Conclusion: The magnetic field dose effects observed in mouse phantoms using a Co-60 beam with 1.5T or 3T fields and a Cs-137 beam with a 3T field compare well with those seen in simulated human treatments with an MRI-linac. These irradiator/magnet combinations are suitable for preclinical studies investigating potential biological effects of delivering radiation therapy in the presence of a magnetic field. Partially funded by Elekta
Additional details
Identifiers
- DOI
- 10.1118/1.4889201;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 41
- Journal Issue
- 6
- Journal Page Range
- p. 436
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46115482
- Subject category
- S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- BIOLOGICAL EFFECTS; CESIUM 137; COBALT 60; LINEAR ACCELERATORS; LUNGS; MICE; MONTE CARLO METHOD; NMR IMAGING; PHANTOMS; PHOTON BEAMS; RADIATION DOSE DISTRIBUTIONS
- Descriptors DEC
- ACCELERATORS; ANIMALS; BEAMS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; CALCULATION METHODS; CESIUM ISOTOPES; COBALT ISOTOPES; DIAGNOSTIC TECHNIQUES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MAMMALS; MINUTES LIVING RADIOISOTOPES; MOCKUP; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; ORGANS; RADIOISOTOPES; RESPIRATORY SYSTEM; RODENTS; STRUCTURAL MODELS; VERTEBRATES; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- (c) 2014 American Association of Physicists in Medicine