Published August 2014 | Version v1
Miscellaneous Open

Calibration of the identiFINDER detector for the iodine measurement in thyroid using the Monte Carlo method

  • 1. Centro de Proteccion e Higiene de las Radiaciones, Calle 20 No. 4113 e/ 41 y 47, Playa, 10600 La Habana (Cuba)

Description

This work is based on the determination of the detection efficiency of 125I and 131I in thyroid of the identiFINDER detector using the Monte Carlo method. The suitability of the calibration method is analyzed, when comparing the results of the direct Monte Carlo method with the corrected, choosing the latter because the differences with the real efficiency stayed below 10%. To simulate the detector their geometric parameters were optimized using a tomographic study, what allowed the uncertainties minimization of the estimates. Finally were obtained the simulations of the detector geometry-point source to find the correction factors to 5 cm, 15 cm and 25 cm, and those corresponding to the detector-simulator arrangement for the method validation and final calculation of the efficiency, demonstrating that in the Monte Carlo method implementation if simulates at a greater distance than the used in the Laboratory measurements an efficiency overestimation can be obtained, while if simulates at a shorter distance this will be underestimated, so should be simulated at the same distance to which will be measured in the reality. Also, is achieved the obtaining of the efficiency curves and minimum detectable activity for the measurement of 131I and 125I. In general is achieved the implementation of the Monte Carlo methodology for the identiFINDER calibration with the purpose of estimating the measured activity of iodine in thyroid. This method represents an ideal way to replace the lack of patterns solutions and simulators assuring the capacities of the Internal Contamination Laboratory of the Centro de Proteccion e Higiene de las Radiaciones are always calibrated for the iodine measurement in thyroid. (author)

Files

45110581.pdf

Files (520.2 kB)

Name Size Download all
md5:34f2e6555c3bf582a2c0477a536a744a
520.2 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Calibracion del detector identiFINDER para la medicion de yodo en tiroides utilizando el metodo Monte Carlo

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Mexico, D. F. (Mexico)
Imprint Pagination
17 p.
Report number
INIS-MX--2791

Conference

Title
14. International Symposium on Solid State Dosimetry
Original Conference Title
14. Conferencia Internacional sobre Dosimetria de Estado Solido
Dates
13-16 Apr 2014
Place
Cusco (Peru)