Calibration of the identiFINDER detector for the iodine measurement in thyroid using the Monte Carlo method
Creators
- 1. Centro de Proteccion e Higiene de las Radiaciones, Calle 20 No. 4113 e/ 41 y 47, Playa, 10600 La Habana (Cuba)
Description
This work is based on the determination of the detection efficiency of 125I and 131I in thyroid of the identiFINDER detector using the Monte Carlo method. The suitability of the calibration method is analyzed, when comparing the results of the direct Monte Carlo method with the corrected, choosing the latter because the differences with the real efficiency stayed below 10%. To simulate the detector their geometric parameters were optimized using a tomographic study, what allowed the uncertainties minimization of the estimates. Finally were obtained the simulations of the detector geometry-point source to find the correction factors to 5 cm, 15 cm and 25 cm, and those corresponding to the detector-simulator arrangement for the method validation and final calculation of the efficiency, demonstrating that in the Monte Carlo method implementation if simulates at a greater distance than the used in the Laboratory measurements an efficiency overestimation can be obtained, while if simulates at a shorter distance this will be underestimated, so should be simulated at the same distance to which will be measured in the reality. Also, is achieved the obtaining of the efficiency curves and minimum detectable activity for the measurement of 131I and 125I. In general is achieved the implementation of the Monte Carlo methodology for the identiFINDER calibration with the purpose of estimating the measured activity of iodine in thyroid. This method represents an ideal way to replace the lack of patterns solutions and simulators assuring the capacities of the Internal Contamination Laboratory of the Centro de Proteccion e Higiene de las Radiaciones are always calibrated for the iodine measurement in thyroid. (author)
Files
45110581.pdf
Files
(520.2 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:34f2e6555c3bf582a2c0477a536a744a
|
520.2 kB | Preview Download |
Additional details
Additional titles
- Original title (Spanish)
- Calibracion del detector identiFINDER para la medicion de yodo en tiroides utilizando el metodo Monte Carlo
Publishing Information
- Publisher
- Sociedad Mexicana de Irradiacion y Dosimetria
- Imprint Place
- Mexico, D. F. (Mexico)
- Imprint Pagination
- 17 p.
- Report number
- INIS-MX--2791
Conference
- Title
- 14. International Symposium on Solid State Dosimetry
- Original Conference Title
- 14. Conferencia Internacional sobre Dosimetria de Estado Solido
- Dates
- 13-16 Apr 2014
- Place
- Cusco (Peru)
INIS
- Country of Publication
- Mexico
- Country of Input or Organization
- Mexico
- INIS RN
- 45110581
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S97: MATHEMATICAL METHODS AND COMPUTING;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; COMPARATIVE EVALUATIONS; COMPUTERIZED SIMULATION; CORRECTIONS; DETECTION; DIAGRAMS; DISTANCE; EFFICIENCY; GEOMETRY; IMPLEMENTATION; IODINE; IODINE 125; IODINE 131; MATHEMATICAL SOLUTIONS; MINIMIZATION; MONTE CARLO METHOD; POINT SOURCES; SIMULATORS; THYROID; VALIDATION
- Descriptors DEC
- ANALOG SYSTEMS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; CALCULATION METHODS; DAYS LIVING RADIOISOTOPES; ELECTRON CAPTURE RADIOISOTOPES; ELEMENTS; ENDOCRINE GLANDS; EVALUATION; FUNCTIONAL MODELS; GLANDS; HALOGENS; INFORMATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IODINE ISOTOPES; ISOTOPES; MATHEMATICS; NONMETALS; NUCLEI; ODD-EVEN NUCLEI; OPTIMIZATION; ORGANS; RADIATION SOURCES; RADIOISOTOPES; SIMULATION; TESTING
Optional Information
- Funding organization
- Organizacion Panamericana de la Salud, Washington D. C. (United States); Universidad Nacional de San Antonio de Abad del Cusco, Cusco (Peru); Instituto Peruano de Energia Nuclear, Lima (Peru); Elekta AB (Sweden); RadPro International GmbH (Germany); ThermoFisher Scientific, Inc. (United States); Varian Medical Systems (United States)