Monte Carlo simulations in multi-detector CT (MDCT) for two PET/CT scanner models using MASH and FASH adult phantoms

  • 1. Department of Physics, Federal University of Sergipe – UFS, São Cristóvão, 49.100-000 (Brazil)
  • 2. Bahia Federal Institute of Education, Science and Technology – IFBA, Vitória da Conquista, 45.100-000 (Brazil)
  • 3. Department of Civil Engineering, Vale do Acarau State University – UVA, Sobral 62.040-730 (Brazil)

Description

The combination of positron emission tomography (PET) and computed tomography (CT) has been extensively used in oncology for diagnosis and staging of tumors, radiotherapy planning and follow-up of patients with cancer, as well as in cardiology and neurology. This study determines by the Monte Carlo method the internal organ dose deposition for computational phantoms created by multidetector CT (MDCT) beams of two PET/CT devices operating with different parameters. The different MDCT beam parameters were largely related to the total filtration that provides a beam energetic change inside the gantry. This parameter was determined experimentally with the Accu-Gold Radcal measurement system. The experimental values of the total filtration were included in the simulations of two MCNPX code scenarios. The absorbed organ doses obtained in MASH and FASH phantoms indicate that bowtie filter geometry and the energy of the X-ray beam have significant influence on the results, although this influence can be compensated by adjusting other variables such as the tube current–time product (mAs) and pitch during PET/CT procedures

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2014.09.036

Additional details

Identifiers

DOI
10.1016/j.nima.2014.09.036;
PII
S0168-9002(14)01064-X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
784
Journal Page Range
p. 524-530
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
15. symposium on radiation measurements and applications
Acronym
SORMA XV
Dates
9-12 Jun 2014
Place
Ann Arbor, MI (United States)