Advantages of magnification in digital phase-contrast mammography using a practical X-ray tube

  • 1. R and D Center, Konica Minolta Medical and Graphic, Inc., 2970 Ishikawa-machi, Hachiouji-shi, Tokyo 192-8505 (Japan)

Description

Phase-contrast imaging with a practical cone-beam X-ray tube has been realized for clinical use in digital mammography using computed radiography (CR). To perform phase-contrast imaging, the X-ray detector must be distanced from an object so that the phase-contrast image achieves magnification; in a mammography unit dedicated to phase-contrast imaging, the magnification ratio is 1.75. When using an X-ray tube with a 0.1-mm focal spot, it appears that the penumbra in magnification blurs both projected images and the phase contrast, which generates an edge effect. However, where the sampling pitch of the CR plate is 43.75 μm, the blur stretches the width of the phase contrast so that unit pixels in the detector can capture it. Note that the width of an ideal phase contrast using an X-ray point source results in a phase contrast too narrow for detection with CR. In addition to phase contrast improving image quality, a re-scaling effect increases image sharpness in magnification. Further, image noise induced by magnification can be reduced during printing to photothermographic dry film by demagnifying the digital output image to the original image size. After demagnification by 1/1.75 from 43.75 μm in image acquisition, a 25-μm pixel size of the output image is obtained so that the spatial resolution matches that of conventional screen-film mammography. In this paper, such technical advantages of magnification in digital phase-contrast mammography are reviewed, and the image quality of phase-contrast images is discussed in light of diagnostic requirements in detecting breast cancer

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ejrad.2008.04.041

Additional details

Identifiers

DOI
10.1016/j.ejrad.2008.04.041;
PII
S0720-048X(08)00282-9;

Publishing Information

Journal Title
European Journal of Radiology
Journal Volume
68
Journal Issue
3,suppl.1
Journal Page Range
p. 69-72
ISSN
0720-048X
CODEN
EJRADR

Conference

Title
5. international workshop medical application of synchrotron radiation 2007
Acronym
MASR2007
Dates
26-30 Aug 2007
Place
Saskatoon (Canada)

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
40044092
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; COMPUTERIZED TOMOGRAPHY; IMAGES; MAMMARY GLANDS; NEOPLASMS; POINT SOURCES; SAMPLING; SPATIAL RESOLUTION; X-RAY TUBES
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DISEASES; ELECTRON TUBES; EQUIPMENT; GLANDS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIATION SOURCES; RADIOLOGY; RESOLUTION; TOMOGRAPHY; X-RAY EQUIPMENT

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.