Low dose MDCT of the wrist-An ex vivo approach
Creators
- 1. Department of Diagnostic Radiology, University Hospital Schleswig-Holstein Campus Kiel, Arnold-Heller Strasse 9, 24105 Kiel (Germany)
- 2. Institute of Medical Informatics and Statistics, University Hospital Schleswig-Holstein Campus Kiel, Brunswiker-Strasse 10, 24105 Kiel (Germany)
- 3. Seminar for Radiation Protection Northern Germany, Christian-Albrechts-University of Kiel, Olshausstrasse 40, 24098 Kiel (Germany)
- 4. Department of Anatomy, Christian-Albrechts-University of Kiel, Otto-Hahn-Platz 8, 24118 Kiel (Germany)
- 5. Siemens Healthcare, Siemens AG, Siemensstrasse 1, 91301 Forchheim (Germany)
Description
The primary objective of this study was to evaluate, if in multidetector computed tomography (MDCT) of the wrist a good image quality can be maintained while radiation dose is substantially reduced. In a second approach one solely parameter change that allows for the best trade-off between dose reduction and image quality should be identified. Twenty wrist specimens were examined with a 16-slice MDCT in different parameter combinations: 120 and 100 kV, 100, 70 and 40 electronic mA s, pitch factor 0.9 and 1.5. Images were reconstructed in four standard planes (slice thickness 1.0 mm, increment 0.5 mm, hard kernel) resulting into a total number of 960 images. Two observers evaluated image quality in a blinded and randomized consensus scheme. Detail quality of corticalis, spongiosa, articular surface and soft tissues was graded according to a four-point scale (1 = excellent, 2 = good, 3 = sufficient, and 4 = poor). The scan protocol with the best trade-off between radiation exposure and image quality had a parameter constellation of 100 kV, 70 electronic mA s (78 effective mA s) and a pitch of 0.9 (DLP 63 mGy cm). This represented a dose reduction of 55%. A solely decrease of voltage lead to a dose reduction of 36% without any loss of image quality. An increase of the pitch factor to 1.5 and a decrease from 70 to 40 mA s caused the most distinct impairment of image quality. In MDCT of the wrist good image quality could be maintained while radiation dose was considerably reduced. A reduction of voltage offers the best result for a solely parameter change.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ejrad.2009.10.024Additional details
Identifiers
- DOI
- 10.1016/j.ejrad.2009.10.024;
- PII
- S0720-048X(09)00588-9;
Publishing Information
- Journal Title
- European Journal of Radiology
- Journal Volume
- 77
- Journal Issue
- 2
- Journal Page Range
- p. 207-214
- ISSN
- 0720-048X
- CODEN
- EJRADR
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43005967
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ARMS; COMPUTERIZED TOMOGRAPHY; IMAGES; RADIATION DOSES; REDUCTION
- Descriptors DEC
- BODY; CHEMICAL REACTIONS; DIAGNOSTIC TECHNIQUES; DOSES; LIMBS; TOMOGRAPHY