Individualized radiation dose control in 256-slice CT coronary angiography (CTCA) in retrospective ECG-triggered helical scans: Using a measure of body size to adjust tube current selection
Creators
- 1. Department of Radiology, Guangdong Academy of Medical Sciences, Guangdong General Hospital, 106 Zhongshan Er Road, Guangzhou 510080 (China)
- 2. CT Clinical Science, Philips Healthcare, Highland Heights, OH (United States)
Description
Purpose: To reduce radiation dose for retrospective ECG-triggered helical 256-slice CTCA by determining an optimal body size index to prospectively adjust tube current. Methods: 102 consecutive patients with suspected CAD underwent retrospective ECG-triggered CTCA using 256-slice CT scanner. Six body size indexes including BMI, nipple level (NL) bust, thoracic anteroposterior diameter at NL, chest circumference (CC) at NL, left main and right coronary artery (RCA) origin level were measured and their correlation with noise was evaluated using linear regression. An equation was developed to use this index to adjust tube current. Additional 102 consecutive patients were scanned with the index-based mA s adjustment. A t-test for independent samples was used to compare radiation dose levels with and without the index-based mA s selection method. Results: Linear regression indicated that CC RCA had the best correlation with noise (R2 = 0.603). Effective radiation dose was reduced from 16.6 ± 0.9 to 9.8 ± 2.7 mSv (p < 0.01), i.e. 40.9% lower dose with the CC RCA-adapted tube current method. The image quality scores indicated no significant difference with and without the size-based mA s selection method. Conclusion: An accessible measure of body size, such as CC RCA, can be used to adapt tube current for individualized radiation dose control.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ejrad.2012.04.017Additional details
Identifiers
- DOI
- 10.1016/j.ejrad.2012.04.017;
- PII
- S0720-048X(12)00192-1;
Publishing Information
- Journal Title
- European Journal of Radiology
- Journal Volume
- 81
- Journal Issue
- 11
- Journal Page Range
- p. 3146-3153
- ISSN
- 0720-048X
- CODEN
- EJRADR
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 45027643
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CHEST; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; CONTROL; CORONARIES; CORRELATIONS; DIAGNOSIS; IMAGES; INDEXES; NOISE; PATIENTS; RADIATION DOSES
- Descriptors DEC
- ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DOCUMENT TYPES; DOSES; EVALUATION; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.