Published August 2007 | Version v1
Miscellaneous

Applying a CPLD for Refurbishment of a Multi-channel Pulse Height Analyzer

Description

Full text: This research applied a CPLD for construction of a 100 MHz, 2048 channel, Wilkinson type analog to digital converter (ADC) circuits for refurbishment of an original multi-channel pulse height analyzer (PHA) ADC. Introduction of the CPLD could reduce the complexity of the circuits, equipment size and also the power consumption while the operation speed was increased. The linearity test of the ADC was found to be excellent with an R2 = 0.9995 and a maximum pulse rate of 48.828 k cps could be converted in this system. Therefore the developed system was appropriate for replacing the original ADC

Availability note (English)

Available in abstract form only, full text entered in this record. Available from Thailand Institute of Nuclear Technology, Bangkok (Thailand)

Additional details

Publishing Information

Imprint Pagination
9 p.
Report number
INIS-TH--162

Conference

Title
10. Conference on Nuclear Science and Technology
Dates
16-17 Aug 2007
Place
Bangkok (Thailand)

INIS

Country of Publication
Thailand
Country of Input or Organization
Thailand
INIS RN
40048204
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference, Numerical Data, Non-conventional Literature
Descriptors DEI
ANALOG-TO-DIGITAL CONVERTERS; ELECTRONIC CIRCUITS; EQUIPMENT; EXPERIMENTAL DATA; HEIGHT; MAXIMUM CREDIBLE ACCIDENT; PULSE ANALYZERS; TOOLS
Descriptors DEC
ACCIDENTS; DATA; DESIGN BASIS ACCIDENTS; DIMENSIONS; ELECTRONIC EQUIPMENT; EQUIPMENT; INFORMATION; NUMERICAL DATA; REACTOR ACCIDENTS

Optional Information

Secondary number(s)
NST--10-ET11