Possibility of semiconductor counters application for internal contamination measurement of whole human body
- 1. Institut za Nuklearne nauke VINCA, Belgrade (Yugoslavia)
Description
The possibility of high resolution semiconductor counters application in 'Vinca' Whole Body Counter for direct beta-gamma internal contamination measurement are discussed, assuming the following relevant characteristics: efficiency, resolution and counter price. A comparison with appropriate characteristics of NaI(Tl) crystal used in 'Vinca' WBC is treated. It is evident that the scintillation counters have the higher detection efficiency, but HPGe counters having much better resolution and recently lowest prices are also acceptable to join the existing NaI(Tl) counters with the HPGe counters in the same time, allow better spectral analyses of the human body activity and additionally more precise estimation of the equivalent doses rate which is generally a essential problem in WBC measurements. (author)
Additional details
Additional titles
- Original title (Serbian)
- Mogucnosti primene poluprovodnickih brojaca za merenje interne kontaminacije celog covecjeg tela
Publishing Information
- Publisher
- Institut za Nuklearne nauke VINCA.
- Imprint Place
- Belgrade (Yugoslavia)
- ISBN
- 86-7306-008-7
- Imprint Title
- Proceedings of XIX Yugoslav symposium of radiation protection, Yugoslav Radiological Protection Association, Belgrade (Yugoslavia), Institut za nuklearne nauke VINCA, Belgrade (Yugoslavia)
- Imprint Pagination
- 406 p.
- Journal Page Range
- p. 137-140.
Conference
- Title
- 19. Yugoslav symposium of radiation protection.
- Original Conference Title
- 19. Jugoslovenski simpozijum za zastitu od zracenja
- Dates
- 18-20 Jun 1997.
- Place
- Golubac (Yugoslavia).
INIS
- Country of Publication
- Yugoslavia
- Country of Input or Organization
- Yugoslavia
- INIS RN
- 28067837
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CONTAMINATION; EFFICIENCY; NAI DETECTORS; RESOLUTION; SEMICONDUCTOR DETECTORS; WHOLE-BODY COUNTERS
- Descriptors DEC
- MEASURING INSTRUMENTS; RADIATION DETECTORS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS
Optional Information
- Notes
- Imprint:7 refs.; 1 figs.; 1 tabs. ;Zbornik radova/ XIX Jugoslovenski simpozijum za zastitu od zracenja/ Jugoslovensko drustvo za zastitu od zracenja; Institut za nuklearne nauke VINCA,Laboratorija za zastitu od zracenja i zastitu sredine 'Zastita'.