Published August 1992
| Version v1
Journal article
The dependence of the energy resolution of gas proportional scintillation counters on the scintillation region to photomultiplier distance
- 1. Physics Dept., Univ. of Coimbra, P-3000, Coimbra (Portugal)
Description
In this paper, the energy resolution for broad X-ray beams of Gas Proportional Scintillation Counters with medium to large size windows, parallel grids and no focusing system is discussed. the variation of the pulse amplitude with the radial distance is calculated for several distances, d (taken from the bottom of the scintillation region to the photomultiplier's photocathode), in the 0 to 50mm range and compared with experimental results. It is shown that for a 2 inch photomultiplier d distances from 10 to 20mm, commonly used in the detector designs, represent a worst case situation
Additional details
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 39
- Journal Issue
- 4
- Journal Page Range
- p. 541-545.
- ISSN
- 0018-9499
- CODEN
- IETNAE
Conference
- Title
- 1991 Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
- Dates
- 2-9 Nov 1991.
- Place
- Santa Fe, NM (United States).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 24047907
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DESIGN; ENERGY SPECTRA; GRIDS; PHOTOCATHODES; PHOTOMULTIPLIERS; PROPORTIONAL COUNTERS; SCINTILLATION COUNTERS; WINDOWS
- Descriptors DEC
- CATHODES; ELECTRODES; MEASURING INSTRUMENTS; OPENINGS; PHOTOTUBES; RADIATION DETECTORS; SPECTRA
Optional Information
- Secondary number(s)
- CONF-911106--.