Published November 1997 | Version v1
Journal article

Isocenter Verification Using Linac-Gram Films Taken with Angiolocalizer : Improved Quality Assurance of Fractionated Stereotactic Radiation Therapy(FSRT)

  • 1. Dept. of Radition Oncology, Samsung Medical Center, Seoul (Korea, Republic of)

Description

With the advances in radiation therapy technology and equipment, the need for more accurate and safer radiation delivery to the target region has been continuously growing. Stereotactic Radiosurgery(SRS) is a good example of 'accuracy' but has a substantial risk of causing severe late neurological damages. Fractionated Stereotactic Radiation Therapy(FSRT) is a modification of SRS enabling conventional fractionation with maintaining accuracy using noninvasive and relocatable frame. Verification of mechanical accuracy in FSRT has been done according to the manufacture's recommendations using RLPP, LTLF, and Depth-helmet. In order to reinforce this, we have developed additional novel verification procedure using Linac-grams with the Angiolocalizer attached on the GTC frame, which are then digitized into the planning software(X-Knife) to generate the three dimensional coordinates for comparison. This method has been successful in such ways that the anatomical landmarks are identifiable on the Linac-gram films and that the serial comparisons of the stereotactic coordinates of the isocenter are possible with more certainty a along the FSRT course than before.

Additional details

Publishing Information

Journal Title
Journal of the Korean Society for Radiotherapeutic Technology
Journal Volume
9
Journal Issue
1
Series
5 refs, 2 figs
Journal Page Range
p. 25-28
ISSN
1598-8449

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
43115857
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
EQUIPMENT; QUALITY ASSURANCE; RADIOTHERAPY; SURGERY; VERIFICATION
Descriptors DEC
MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY