Published November 1989 | Version v1
Journal article

Online brain attenuation correction in PET: Towards a fully automated data handling in a clinical environment

  • 1. Louvain Univ., Louvain-la-Neuve (Belgium). Lab. of Positron Emission Tomography

Description

We have improved the calculation of the brain attenuation correction in positron emission tomography (PET) and set up a procedure which allows the clinician to get a fully corrected image in a single reconstruction step, without human intervention. By using a general object description scheme based on polygonal contour trees we are able to calculate the attenuation correction for brain tissue, bone and head holder. The head contour is generated from the emission sinogram. On a set of 15 adult patients, the emission values obtained using this calculated attenuation compare favorably with those obtained with an attenuation resulting from a transmission measurement. Residual discrepancies are attributed to incomplete scatter compensation between emission and transmission. The robustness of the algorithm has been tested on more than 100 brain fluorodeoxyglucose (18FDG) studies in adult patients, including pathological cases. Its applicability for 18FDG studies in children and for other tracers such as water (H215O) and fluoroethylspiperone (18FESP) is also presented. (orig.)

Additional details

Publishing Information

Journal Title
European Journal of Nuclear Medicine
Journal Volume
15
Journal Issue
11
Series
Eur. J. Nucl. Med.
Journal Page Range
712-718
ISSN
0340-6997
CODEN
EJNMD