Simulation of Dose to Surrounding Normal Structures in Tangential Breast Radiotherapy Due to Setup Error

  • 1. Department of Radiology, All India Institute of Medical Sciences, New Delhi (India)
  • 2. Department of Nuclear Medicine, All India Institute of Medical Sciences, New Delhi (India)
  • 3. Department of Radiotherapy and Oncology, Institute Rotary Cancer Hospital, All India Institute of Medical Sciences, New Delhi (India)

Description

Setup error plays a significant role in the final treatment outcome in radiotherapy. The effect of setup error on the planning target volume (PTV) and surrounding critical structures has been studied and the maximum allowed tolerance in setup error with minimal complications to the surrounding critical structure and acceptable tumor control probability is determined. Twelve patients were selected for this study after breast conservation surgery, wherein 8 patients were right-sided and 4 were left-sided breast. Tangential fields were placed on the 3-dimensional-computed tomography (3D-CT) dataset by isocentric technique and the dose to the PTV, ipsilateral lung (IL), contralateral lung (CLL), contralateral breast (CLB), heart, and liver were then computed from dose-volume histograms (DVHs). The planning isocenter was shifted for 3 and 10 mm in all 3 directions (X, Y, Z) to simulate the setup error encountered during treatment. Dosimetric studies were performed for each patient for PTV according to ICRU 50 guidelines: mean doses to PTV, IL, CLL, heart, CLB, liver, and percentage of lung volume that received a dose of 20 Gy or more (V20); percentage of heart volume that received a dose of 30 Gy or more (V30); and volume of liver that received a dose of 50 Gy or more (V50) were calculated for all of the above-mentioned isocenter shifts and compared to the results with zero isocenter shift. Simulation of different isocenter shifts in all 3 directions showed that the isocentric shifts along the posterior direction had a very significant effect on the dose to the heart, IL, CLL, and CLB, which was followed by the lateral direction. The setup error in isocenter should be strictly kept below 3 mm. The study shows that isocenter verification in the case of tangential fields should be performed to reduce future complications to adjacent normal tissues

Availability note (English)

Available from http://dx.doi.org/10.1016/j.meddos.2007.03.003

Additional details

Identifiers

DOI
10.1016/j.meddos.2007.03.003;
PII
S0958-3947(07)00057-X;

Publishing Information

Journal Title
Medical Dosimetry
Journal Volume
33
Journal Issue
1
Journal Page Range
p. 81-85
ISSN
0958-3947

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.