Published September 1998 | Version v1
Journal article

Evaluation of three-dimensional gadolinium-enhanced MR angiography using the timing monitoring function of contrast material (Smart Prep technique)

  • 1. Nippon Medical School, Tokyo (Japan). Main Hospital

Description

The Smart Prep technique for gadolinium-enhanced three-dimensional MR angiography (3D-MRA) was evaluated in clinical practice. By monitoring signal intensity in the region of interest (tracking volume) in the target vessel, start timing after contrast injection can be optimized using the Smart Prep technique. Successful triggering was obtained in the chest, abdomen, and pelvic areas in about 80% of the cases in this study. Failures with this technique were mainly due to changes in tracking volume caused by patient motion and respiration. We noted that the scan started earlier than expected in the thoracic aorta when part of the heart or pulmonary artery was included in the tracking volume. Thus, care must be taken in defining the size and location of the tracking volume in gadolinium-enhanced 3D-MRA using the Smart Prep technique. (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
54
Journal Issue
9
Journal Page Range
p. 1155-1163
ISSN