Published July 1998 | Version v1
Journal article

On atomization of uranium in plutonium matrix using ETA-AAS

  • 1. Radiochemistry Division, Bhabha Atomic Research Centre, Trombay, Mumbai (India)

Description

Atomization processes for uranium in aqueous media and in the presence of a plutonium matrix have been studied and a chemical mechanism is proposed. These studies have been utilized for the determination of uranium in plutonium by Electrothermal Atomization- Atomic Absorption Spectrometry (ETA-AAS) within the constraints of its stable carbide forming tendency and complexity caused by formation of plutonium suboxide at high temperatures. In spite of these limitations the analytical range obtained for determination of uranium is 2.5-100 ng with a sample aliquot of 5 μL containing 5 mg/mL plutonium concentration. The precision of the method is evaluated as 9% RSD. (orig.)

Additional details

Publishing Information

Journal Title
Fresenius' Journal of Analytical Chemistry
Journal Volume
361
Journal Issue
5
Journal Page Range
p. 429-432
ISSN
0937-0633
CODEN
FJACES

Optional Information

Notes
With 1 fig., 3 tabs., 10 refs.