Nuclear Regulatory Authority low energy germanium detection system: performance for the uranium individual monitoring
- 1. Autoridad Regulatoria Nuclear, Avda. del Libertador 8250, (C1429BNP) Ciudad Autonoma de Buenos Aires (Argentina)
- 2. Sociedad Argentina de Radioproteccion, Avda. del Libertador 8250, (C1429BNP) Ciudad Autonoma de Buenos Aires (Argentina)
Description
The lung counter facility of the Nuclear Regulatory Authority (ARN) is presented. A calibration was carried out using the Lawrence Livermore National Laboratory (LLNL) phantom. This phantom is provided with a pair of lungs and lymph nodes containing uranium homogeneously distributed and a set of four overlay plates covering a chest wall thickness (CWT) ranging from 1.638 to 3.871 cm. Individual organ calibration factors were acquired for 235U photo-peaks energies and for each effective chest thickness. Using these factors, a collection of theoretical fitting curves were found. A counting efficiency formulae and a curve for simultaneously active lymph nodes and lung was obtained and checked through measures. Background measurements of the chamber with and without volunteer persons were performed in order to obtain the detection limits (DL) of the system. As this task involves the knowledge of the volunteers CWTs, these magnitudes were determined through formulae selected from the literature taking into account the detection system characteristics. The deviation in the CWT assigned to an individual, generated by applying different equations, produces variations up to 33% in the estimations of the incorporated activity and DL. An analysis of the changes in efficiencies as consequences of the detectors locations and CWT was also performed. This reveals that the DL of the camera (detectors, shield and blank phantom) is between 2.7 and 6.4 Bq of 235U, which implies 4.9 and 11.5 mg lung burden of natural uranium. An estimation of the minimum detectable intake performed with the DL considering blank persons shows that a system with the characteristics described is only adequate for non-routine individual monitoring. (authors)
Availability note (English)
Available from doi:Additional details
Identifiers
- DOI
- 10.1093/rpd/ncm482;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 127
- Journal Issue
- 1-4
- Journal Page Range
- p. 287-292
- ISSN
- 0144-8420
Conference
- Title
- Internal dosimetry of radionuclides - Occupational, Public and Medical Exposure
- Dates
- 2-5 Oct 2006
- Place
- Montpellier (France)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 39096334
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; CHEST; EFFICIENCY; GERMANIUM; LUNGS; LYMPH NODES; NATURAL URANIUM; PERFORMANCE; PHANTOMS; RADIATION MONITORING; SHIELDS; THICKNESS; URANIUM 235
- Descriptors DEC
- ACTINIDE NUCLEI; ACTINIDES; ALPHA DECAY RADIOISOTOPES; BODY; DIMENSIONS; ELEMENTS; EVEN-ODD NUCLEI; HEAVY NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LYMPHATIC SYSTEM; METALS; MINUTES LIVING RADIOISOTOPES; MOCKUP; MONITORING; NUCLEI; ORGANS; RADIOISOTOPES; RESPIRATORY SYSTEM; SPONTANEOUS FISSION RADIOISOTOPES; STRUCTURAL MODELS; URANIUM; URANIUM ISOTOPES; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 14 refs