Ventricular fibrillation time constant for swine
Creators
- 1. Department of Electrical and Computer Engineering, University of Wisconsin, Madison, WI 53706 (United States)
- 2. Department of Biomedical Engineering, University of Wisconsin, 1550 Engineering Drive, Madison, WI 53706 (United States)
- 3. Department of Surgery, University of Wisconsin, Madison, WI 53792 (United States)
- 4. Department of Nutritional Sciences, University of Wisconsin, Madison, WI 53706 (United States)
- 5. Departments of Pathobiology, SVM, and Animal Sciences, CALS, University of Wisconsin, Madison, WI 53706 (United States)
Description
The strength–duration curve for cardiac excitation can be modeled by a parallel resistor–capacitor circuit that has a time constant. Experiments on six pigs were performed by delivering current from the X26 Taser dart at a distance from the heart to cause ventricular fibrillation (VF). The X26 Taser is an electromuscular incapacitation device (EMD), which generates about 50 kV and delivers a pulse train of about 15–19 pulses s−1 with a pulse duration of about 150 µs and peak current about 2 A. Similarly a continuous 60 Hz alternating current of the amplitude required to cause VF was delivered from the same distance. The average current and duration of the current pulse were estimated in both sets of experiments. The strength–duration equation was solved to yield an average time constant of 2.87 ms ± 1.90 (SD). Results obtained may help in the development of safety standards for future electromuscular incapacitation devices (EMDs) without requiring additional animal tests
Availability note (English)
Available from http://dx.doi.org/10.1088/0967-3334/29/10/005Additional details
Identifiers
- DOI
- 10.1088/0967-3334/29/10/005;
- PII
- S0967-3334(08)76553-5;
Publishing Information
- Journal Title
- Physiological Measurement (Print)
- Journal Volume
- 29
- Journal Issue
- 10
- Journal Page Range
- p. 1209-1219
- ISSN
- 0967-3334
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44127320
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ALTERNATING CURRENT; EQUIPMENT; EXCITATION; HEART; PULSES; SAFETY STANDARDS; SWINE; TESTING
- Descriptors DEC
- ANIMALS; BODY; CARDIOVASCULAR SYSTEM; CURRENTS; DOMESTIC ANIMALS; ELECTRIC CURRENTS; ENERGY-LEVEL TRANSITIONS; MAMMALS; ORGANS; STANDARDS; VERTEBRATES