A Study on the Factors of Spatial Scattered Ray Occurrence in the X-ray Radiography Room
Creators
- 1. Dept. of Radiological Science, Gimcheon College, Gimcheon (Korea, Republic of)
- 2. Dept. of Radiation, Ray Laboratory, Seoul (Korea, Republic of)
Description
In this study, we measured the dose distribution of scattered ray in X-ray radiography room using an ion chamber and examined the dependency of scattered ray content on the scattered ray source and exposure condition. To study the factors of scattered ray occurrence in the acryl phantom, we measured the change in the scatted ray content according to the X-ray tube voltage (40-140 kV) and the field size (10 x 10 cm2, 20 x 20 cm2, 35 x 35 cm2). For the 35 x 35 cm2 field size, the side-scattering rate ranged from 3.1% to 14.5%. The scattered ray contributions of the phantom, collimator, X-ray tube and wall were also measured. The scattered ray contribution of the phantom was higher than 95.4% for the entire tube voltage, and those of the collimator, X-ray tube and wall were 2.6%, 1.3% and 0.7%, respectively.
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Society of Radiological Technology
- Journal Volume
- 32
- Journal Issue
- 4
- Series
- 8 refs, 8 figs, 1 tab
- Journal Page Range
- p. 393-399
- ISSN
- 1226-2854
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 44015475
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COLLIMATORS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; SPATIAL DISTRIBUTION; TUBES
- Descriptors DEC
- DISTRIBUTION; DOSES