In this study, we measured the dose distribution of scattered ray in X-ray radiography room using an ion chamber and examined the dependency of scattered ray content on the scattered ray source and exposure condition. To study the factors of scattered ray occurrence in the acryl phantom, we measured the change in the scatted ray content according to the X-ray tube voltage (40-140 kV) and the field size (10 x 10 cm2, 20 x 20 cm2, 35 x 35 cm2). For the 35 x 35 cm2 field size, the side-scattering rate ranged from 3.1% to 14.5%. The scattered ray contributions of the phantom, collimator, X-ray tube and wall were also measured. The scattered ray contribution of the phantom was higher than 95.4% for the entire tube voltage, and those of the collimator, X-ray tube and wall were 2.6%, 1.3% and 0.7%, respectively.

Additional details

Publishing Information

Journal Title
Journal of the Korean Society of Radiological Technology
Journal Volume
32
Journal Issue
4
Series
8 refs, 8 figs, 1 tab
Journal Page Range
p. 393-399
ISSN
1226-2854

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
44015475
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COLLIMATORS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; SPATIAL DISTRIBUTION; TUBES
Descriptors DEC